MultiSystem 4010
Operating Instructions
© Hydrotechnik GmbH • All rights reserved
Rev. 1.5 / 120816 english
Page 42 of 45
Baud rate CAN select the data transmission speed of the CAN bus; this must be identic for all
participants, otherwise communication will not be possible
CAN Power select whether a supply power shall be emitted via the CAN interface
CANopen device highlight this line and press
to send the start command into the CAN bus
6.10.2. Filter settings
Highlight the item „Filter settings“ in the device menu and press :
Here you may use several software filters:
Filter for analog input channels (AD-channels)
The analog input channels are scanned with 1 ms. With the software filter mean values can be
calculated from several measured values to smooth the input signal. You may select between “1 ms”
(no smoothing) and “2/4/8/16 ms” (mean value caculation from 2/4/8/18 measured values). At “16 ms”
the instrument will deliver a new measured value each 16 ms.
Filter for frequency input channels
Information
The parameters for the combined input channel Ch4 will
only be displayed if it is programmed as frequency input
channel.
You may define two parameters for each frequency input channel:
Gate time a new measured value will not be accepted until the gate time expired; enter a
multiple og 10 ms
Lowest frequency select the lowest frequency that shall be measured; this influences the behavior of
the instrument, when the measured value „Zero“ is displayed
Example a lowest frequency of „0.25 Hz“ is defined and the measured object (e.g. a turbine
flow meter) stops running; this would have to be displayed as a measured value of
„0.00“; since the instrument is requested to measure down to 0.25 Hz, it will wait
four seconds for a new measured value, until “0.00” is displayed
In this case you should use a lowest frequency of 10 or 100 Hz, then the stopping
of the turbine will be displayed after 0.1 or 0.01 seconds.
Confirm the settings with
.