MultiSystem 4010
Operating Instructions
© Hydrotechnik GmbH • All rights reserved
Rev. 1.5 / 120816 english
Page 27 of 45
Highlight an item and confirm with
. Further options are displayed due to the selected occupation.
Settings at the functions “ChX-ChY” and „POWER”
Define these options here:
Meas. variable enter the variable resulting from the calculation
Index variable enter the index number of the variable if you have several identic
Units enter the name of the units resulting from the calculation
Name enter a name for the channel
Align. diff. function to compensate the measuring difference between two sensors;
connect both sensors and put an identic load on them (e.g. 0 bar); then
highlight this function and press
; the instrument determins both
measured values and uses a difference for the compensation during the
following measuring
Press
to save the channel settings.
Settings at the function „Chx/dt“
You may use this function e.g. to calculate speed from a measured way. You may define a threshold
value (“Level”) to prevent the calculation of absurd values for small way measurements. This assures
that a certain way has to be measured before a calculation is executed. The gate time assures for a
defined level that a calculation is executed after a certain time. This is the only way to assure the
calculation of a speed “0”.
Define these options:
Meas. variable enter the variable resulting from the calculation
Index variable enter the index number of the variable if you have several identic
Units enter the name of the units resulting from the calculation
Name enter a name for the channel
Level value for that the basic measured value must change, before a
calculation is executed