6.4. Configure Channels
Highlight „Channels“ in the main menu and press :
You can see a list of all twelve channels of the instrument. The first five are the physical input
channels (see section 3.2 on page 6), C6 and C7 are the trigger input resp. output. The other
channels can be used for calculations or may be occupied with digital input signals (e.g. CAN, option).
Use the arrow keys to highlight a channels and press
to open it.
6.4.1. Analog Input Channels (C1 … C3)
Highlight one of the channels C1 … C3 in the channel menu and press :
Use the arrow keys to highlight a menu item and press
to edit the displayed value / setting. The
menu items have these functions:
Meas. Variable type and units of the measurand of the sensor connected to this channel
Index Variable identic variables are indexed automatically
Name enter the desired name of the channel
Signal type select the signal type of the connected sensor (0/4-20 mA or 0/2-10 V)
Measuring range enter the start and end values of the measuring range
Zero point function to execute a zero point equalisation of the sensor; connect the
sensor with a load of „0“ (e.g. no pressure), highlight „Zero point“ and
press
; press to start the equalisation, then the compensation
value will be displayed
Linearisation here you may enter (or select) a linearisation table to compensate the
linearity error of the sensor (see below)
Press
after finishing the inputs to save all settings.