Manual the output can be switched using the parameter „Condition“
If „Meas. channel“ is selected, further parameters will be displayed:
Source select the measuring channel that shall be supervised for the incident that
shall switch the output
Condition select either „greater“ or „lower“; the output will be switched if the value at
the selected channel exceeds / falls below the defined value
Value enter the value for the condition
In the image channel p1 is supervised, the output will be switched if the measured value exceeds 120.
Simultaneous recording on connected instruments
You need a special cable (order N
o
8824-F2-00.50) to connect two MultiSystem 4010, or the
MultiXtend Trigger (order N
o
316A-A0-00.50) to connect up to five instruments. Then it becomes
possible to start build up a master-slave connection and start a recording on all instruments
simultaneously.
At the Master Instrument you should program a recording that is started by a trigger event. Then
select “Trigger event” as operation mode of the trigger output. As soon as the trigger event happens,
the output will be switched.
At the Slave Instruments you should program a recording that is started by the trigger event
„Channel E1 – ON”. Then the recording will be started as soon as the trigger input receives the signal
from the trigger output of the master instrument.
More information on the setup of trigger recordings is contained in section 6.6 on page 33.
6.4.6. Free channels (C8 … C12)
These channels can be used for calculations or as CAN channels (option).
Press
and select one of the available occupations:
UNDEF. channel is not used
Ch1-Ch2 calculation of the difference of the measured values from channel 1 and 2
Ch3-Ch4 calculation of the difference of the measured values from channel 3 and 4
Ch4-Ch5 calculation of the difference of the measured values from channel 4 and 5
Ch1/dt calculates the first derivation of the measured values from channel 1
Ch5/dt calculates the first derivation of the measured values from channel 5
POWER uses the formula „Ch1 x Ch5 / 600“ to calculate the hydraulic power; pressure is
measured on channel 1, volume flow rate on channel 5
CAN a signal from a connected CAN bus (option) is assigned to the channel
Multimeter the signal from a Multimeter connected to the instrument is assigned to this
channel